Bring tests, specimens, buffer, and/or controls to room temperature (15-30°C) before use.
1. Remove the test from its sealed pouch, and place it on a clean, level surface. Label the device with patient or control identification. For best results, the assay should be performed within one hour.
2. Using the disposable pipette provided in the kit, add 2 drops of serum or plasma to the specimen well (S) of the device, then start the timer.
OR
Add 3 drops of whole blood specimen to the specimen well (S) of the device with the provided disposable pipette, then add 1 drop of buffer and start the timer.
OR
Allow 3 hanging drops of fingerstick whole blood specimen to fall into the center of the specimen well (S) on the device, then add 1 drop of buffer and start the timer.
Avoid trapping air bubbles in the specimen well (S), and do not add any solution to the result area.
As the test begins to work, color will migrate across the membrane.
3. Wait for the colored band(s) to appear. The result should be read at 10 minutes. Do not interpret the result after 20 minutes. |